Citação | SILVA, Pedro Henrique Reis da et al. An innovative, simple, fast, and less toxic high-performance liquid chromatographic method for determination of prednisone in capsules. Braz. J. Pharm. Sci, v. 53, n. 2, e16064, 2017. | pt_BR |
ISSN | 2175-9790 | pt_BR |
URI | https://www.arca.fiocruz.br/handle/icict/27155 | |
Idioma | eng | pt_BR |
Editor | Universidade de São Paulo | pt_BR |
Direito Autoral | open access | pt_BR |
Palavras-chave | Prednisona | pt_BR |
Palavras-chave | capsulas | pt_BR |
Palavras-chave | Cromatografia líquida de alta eficiência / validação | pt_BR |
Palavras-chave | Avaliação da linearidade | pt_BR |
Título | An innovative, simple, fast, and less toxic high-performance liquid chromatographic method for determination of prednisone in capsules | pt_BR |
Tipo do documento | Article | pt_BR |
DOI | 10.1590/s2175-97902017000216064 | |
Resumo em Inglês | Prednisone is an anti-inflammatory steroid drug widely used in clinical practice. However, no high-performance liquid chromatographic (HPLC) method has been described in the literature for the determination of prednisone in capsules until now. Thus, an HPLC method was developed using a C18 (250x4.0, 5 µm) column, with methanol:water (70:30) as mobile phase at a flow rate of 1 mL/min and detection at 240 nm. The developed method was validated following current Brazilian legislation. Additionally, linearity was assessed by evaluating the assumptions of normality, homoscedasticity, and independency of residuals, and the fit to the linear model. The method showed linearity (r2>0.99) over the range of 14.0-26.0 µg/mL, selectivity, precision (RSD<2.0%), robustness, and accuracy (average recovery of 100.05%). The chromatographic procedure was applied for assay and uniformity content determination of three different batches of prednisone capsules, showing to be suitable for their quality control. | pt_BR |
Afiliação | Universidade Federal de Minas Gerais. Faculdade de Farmácia. Departamento de Produtos Farmaceuticos. Belo Horizonte, MG, Brazil | pt_BR |
Afiliação | Fundação Oswaldo Cruz. Instituto René Rachou. Belo Horizonte, MG, Brazil | pt_BR |
Afiliação | Universidade Federal de Minas Gerais. Faculdade de Farmácia. Departamento de Produtos Farmaceuticos. Belo Horizonte, MG, Brazil | pt_BR |
Afiliação | Universidade Federal de Minas Gerais. Faculdade de Farmácia. Departamento de Produtos Farmaceuticos. Belo Horizonte, MG, Brazil | pt_BR |
Palavras-chave em inglês | Prednisone | pt_BR |
Palavras-chave em inglês | Capsules | pt_BR |
Palavras-chave em inglês | High performance liquid chromatography/validation | pt_BR |
Palavras-chave em inglês | Linearity assessment | pt_BR |