Abstract: Nos últimos anos, as farmacopéias e principais agências regulatórias para produtos farmacêuticos e biofarmacêuticos exigem cada vez mais em suas monografias a aplicação do método para detecção de endotoxinas bacterianas pelo lisado de amebócitos de Limulus(LAL) no controle de pirogênio de produtos terminados parenterais. O objetivo do presente estudo foi padronizar e validar o ensaio de LAL para o biofármaco alfainterferona 2b humana recombinante, produzido em E. coli. Foram empregados os métodos Gel-Clot e Cinético-Cromogênico (semiquantitativo e quantitativo, respectivamente). Para o método Cinético-Cromogênico, a máxima diluição válida (MDV) foi calculada para cada apresentação com base na concentração do produto e a sensibilidade do teste (3 MUI, MDV: 1:3.888 e 10 MUI, MVD: 1:12.962). Diluições seriadas abaixo da MDV foram avaliadas emtriplicata para a verificação dos interferentes, sendo eleitas as diluições de 1:80 e1:100 que exibiram a melhor recuperação no controle positivo do produto. Para oensaio Gel-Clot na apresentação de 3 MUI (MDV: 1:17) foi estipulada a diluição de 1:8para a validação dos testes de rotina. Durante a validação dos ensaios, foi utilizada uma curva-padrão de endotoxina nas concentrações de 0,005 – 50 EU/ml e avaliados o seucoeficiente de correlação (R) e o desvio-padrão relativo. Os critérios de desempenho do método cinético (linearidade, especificidade, exatidão, repetibilidade, reprodutibilidade) foram atendidos de acordo com os parâmetros farmacopéicos e regulatórios (ANVISA e FDA), garantindo desta forma a robustez necessária para que o método de LAL possa ser incluído nos ensaios de rotina do Laboratório de Controle de Qualidade.
Abstract: In recent years, the Pharmacopean, technical reports and international guides for pharmaceuticals products requires each time more intheir monographies the application of Limulus Ambocytes Lysateendotoxins assay (LAL) for release and pyrogen control in parenteral finished products. The objective of thisstudy was the standardization and validation of the LAL test for the human recombinant biopharmaceutical interferon alpha 2b, produced in E. coli, were used Gel-Clot and Chromogenics Kinetic methods (semi-quantitative and quantitative). In the case of Chromogenics method test, the maximum valid dilution (MDV) was calculated for each presentation based on the concentration of product (3 MUI, MDV: 1:3,888 and 10 MUI, MVD: 1:12,962), serial dilutions under the MDV (1:80) were evaluated in triplicate to detect interferences. For the gel-clot test for the 3 MUI presentation (MDV: 1:17) 1:8 dilution was setfor interferences detection test. For tests’ validation, several dilutions were performedusing standard endotoxin concentrations in 0,005 to 50 EU / ml to confirm the criteria for the methods performance (linearity, specificity, accuracy, reply, reproducibility). The results of validation showed that all pharmacopeia and regulatory parameters (ANVISA) studied, are compatible with the MDV used for each studied presentation of the human recombinant biopharmaceuticals interferon alpha 2b and may be used in quality control.
Keywords: Limulus Amebocyte Lysate
Alfainterferona 2b Humana Recombinante
Chemistry, Pharmaceutical
Quality Control
keywords: Limulus Amebocyte Lysate
Alfainterferona 2b Humana Recombinante
Química Farmacêutica
Controle de Qualidade
DeCS: Química Farmacêutica
Controle de Qualidade
Issue Date: 2009
Publisher: Instituto de Tecnologia em Imunobiológicos.
Citation: BRUM, Ricardo Cristiano de Souza. Padronização e validação da técnica do Limulus Amebocyte Lysate (LAL) semi-quantitativa e quantitativa para o biofármaco Alfainterferona 2B humana recombinante. 2009. 115 f. Dissertação (Mestrado Profissional em Tecnologia de Imunobiológicos) – Instituto de Tecnologia em Imunobiológicos, em parceria com o Instituto Oswaldo Cruz, Fundação Oswaldo Cruz, Rio de Janeiro, 2009.
Date of defense: 2009
Place of defense: Rio de Janeiro / RJ
Department: Departamento de Controle de Qualidade
Defense institution: Fundação Oswaldo Cruz. Instituto de Tecnologia em Imunobiológicos.
Program: Mestrado Profissional em Tecnologia de Imunobiológicos.
Copyright: open access
Appears in Collections:Biomanguinhos - Dissertações de Mestrado

Files in This Item:
File Description SizeFormat 
ricardo-cristiano-de-souza-brum.pdfDocumento principal1.7 MBAdobe PDFThumbnail
View/Open



FacebookTwitterDeliciousLinkedInGoogle BookmarksBibTex Format mendeley Endnote DiggMySpace

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.