Author | Lemgurber, Ana Paula Mello | |
Author | Presgrave, Octavio Augusto França | |
Author | Presgrave, Rosaura de Farias | |
Author | Alves, Eloisa Nunes | |
Author | Silva, Ronald Santos | |
Author | Lopes, Izabela Gimenes | |
Author | Freitas, João Carlos Borges Rolim de | |
Author | Silva, Cristiane Caldeira da | |
Access date | 2014-12-17T16:53:50Z | |
Available date | 2014-12-17T16:53:50Z | |
Document date | 2011 | |
Citation | LEMGRUBER, A. P. M. et al. Influência do erro de micropipetas na dosagem espectrofotométrica do ensaio de endotoxina bacteriana (LAL cromogênico) aplicável a produtos sujeitos à Vigilância Sanitária. Revista do Instituto Adolfo Lutz, São Paulo, v. 70, n. 3, p. 368-372, 2011. | pt_BR |
ISSN | 0073-9855 | |
URI | https://www.arca.fiocruz.br/handle/icict/9202 | |
Abstract in Portuguese | Segundo a norma ISO 17.025, os instrumentos usados em experimento devem ser calibrados. No caso de micropipetas não há limite máximo de erro definido, sendo difícil estipular até que ponto uma micropipeta pode continuar a ser usada. O objetivo do trabalho foi o de determinar a porcentagem de erro no volume da micropipeta que não interfira nos resultados das diferentes etapas do ensaio toxicológico Lisado de Amebócitos de Limulus (LAL). Os percentuais de erro foram obtidos do levantamento de 44 certificados de micropipetas. Foram utilizados uma pipeta monocanal calibrada, de volume variável e kit LAL Cromogênio QCL-1000 (Lonza). A curva de endotoxina foi construída com as concentrações de 0,1; 0,25; 0,5 e 1 UE/mL. Os percentuais de erro de ±2%, ±4% e ±10% foram comparados com a concentração de referência (0,5 UE/mL), e analisados pelo Student t-test (p<0,05). A análise estatística mostrou que, para a endotoxina padrão e para o substrato do LAL, erro acima de 2% interfere nos resultados, cuja correção deve ser feita pelo ajuste do volume na pipeta. Nenhum erro testado causou interferência estatisticamente significativa na reação da cor do substrato. Micropipetas com erro superior a ± 2% devem ser corrigidas | pt_BR |
Language | por | pt_BR |
Publisher | Instituto Adolfo Lutz | pt_BR |
Rights | open access | pt_BR |
Subject in Portuguese | ISO 17.025 | pt_BR |
Subject in Portuguese | Incerteza de Medição | pt_BR |
Subject in Portuguese | Teste de LAL | pt_BR |
Subject in Portuguese | Erros em Micropipetas | pt_BR |
Title | Influência do erro de micropipetas na dosagem espectrofotométrica do ensaio de endotoxina bacteriana (LAL cromogênico) aplicável a produtos sujeitos à Vigilância Sanitária | pt_BR |
Alternative title | Effect of micropipette inaccuracy on performance of spectrophotometric assay for bacteria endotoxin dosage (chromogenic LAL) employed in products subjected to sanitary surveillance | en |
Type | Article | pt_BR |
Abstract | The ISO 17025 Guidelines states that all instruments used in performing experiments must be calibrated. For micropipettes no maximum error limit has established, which makes difficult to determine as far as a micropipette can be used without interfering on the final results. The aim of this study was to determine the percentage of errors which could be accepted in the micropipettes volume without interfering on results from different steps of LAL toxicological assay. The error percentages were determined based on the information from 44 certificates of micropipettes survey. A monochannel variable volume calibrated micropipette (Eppendorf) and QCL Chromogenic LAL kit (Lonza) were employed. Endotoxin curve was plotted using 0.1, 0.25, 0.5 and 1 EU/mL concentrations. The error percentages of ±2%, ±4% and ±10% were compared with the reference concentration (0.5 EU/mL). Results were analyzed by Student’s t-Test in relation to the reference concentration (p<0.05). Statistical analysis showed that the error of 2% did not interfere on standard endotoxin and LAL substrate results, but errors above this percentage caused altera-tions on results. Micropipette errors did not interfere on the reaction of substrate color. Errors higher than ±2% should be corrected by adjusting the micropipette volume. | en |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Affilliation | Fundação Oswaldo Cruz. Instituto Nacional de Controle de Qualidade em Saúde. Departamento de Farmacologia e Toxicologia. Rio de Janeiro, RJ, Brasil. | pt_BR |
Subject | ISO 17.025 | en |
Subject | Measurement Unsureness | en |
Subject | LAL test | en |
Subject | Micropipette Error | en |
DeCS | Normas Técnicas | pt_BR |
DeCS | Equipamentos de Medição | pt_BR |
DeCS | Espectrofotometria | pt_BR |
DeCS | Endotoxinas | pt_BR |